물 자연 그리고 사람 - 물로 더 행복한 세상을 만들어가겠습니다.
HOMER&D성과논문실적

논문실적

Simplification and Application of Large Volume Sampler for Routine Monitoring 게시글의 제목, 학술지명, 저자, 발행일, 작성내용을 보여줌
Simplification and Application of Large Volume Sampler for Routine Monitoring
학술지명 한국환경분석학회 저자 백경희,이재희,김정희,최재원,문부식,윤병석
발표일 2007-05-17

The Large volume extraction system based on polyurethane foam (PUF) for the field extraction was simplified for routine monitoring of polychlorinated dibenzo-p-dioxins (PCDDs) and dibenzofurans (PCDFs) in drinking water treatment plant. After pre-test for the whole system, validation for simplified large volume sampler (SLV) was performed. The flow rate was tested triplicates between 1 L/min, 2 L/min, and 3 L/min, respectively. Accuracy as a recoveries of the absorption parts against 1~3 L/min were 86-96 %, 80-110 %, and 93-107 %, respectively. Precision as a relative standard deviation (RSD) for the tested flow rates was 5.1 %, 15.3 %, and 7.5 %, respectively. This indicates that similar results are expected between 1-3 L/min in field extraction using PUF-type large volume sampler. SLV system showed that reduced size/maintenance, extraction time and solvent wastes compare to the other LV system adopting the same PUF plugs.

목록